thumb|A Helmholtz coil

thumb|255px|Helmholtz coil schematic drawing

A Helmholtz coil is a device for producing a region of nearly uniform magnetic field, named after the German physicist Hermann von Helmholtz. It consists of two electromagnets on the same axis, carrying an equal electric current in the same direction. Besides creating magnetic fields, Helmholtz coils are also used in scientific apparatus to cancel external magnetic fields, such as the Earth's magnetic field.

thumb|upright=1.2|A beam of cathode rays in a vacuum tube bent into a circle by a Helmholtz coil

Description

A Helmholtz pair consists of two identical circular magnetic coils that are placed symmetrically along a common axis, one on each side of the experimental area, and separated by a distance <math>h</math> equal to the radius <math>R</math> of the coil. Each coil carries an equal electric current in the same direction.

Setting <math>h=R</math>, which is what defines a Helmholtz pair, minimizes the nonuniformity of the field at the center of the coils, in the sense of setting <math>\partial^{2}B/\partial x^{2} = 0</math> (meaning that the first nonzero derivative is <math>\partial^{4}B/\partial x^{4}</math> as explained below), but leaves about 7% variation in field strength between the center and the planes of the coils.

A slightly larger value of <math>h</math> reduces the difference in field between the center and the planes of the coils, at the expense of worsening the field's uniformity in the region near the center, as measured by <math>\partial^{2}B/\partial x^{2}</math>.

In some applications, a Helmholtz coil is used to cancel out the Earth's magnetic field, producing a region with a magnetic field intensity much closer to zero.

Mathematics

[[File:VFPt helmholtz coil thumb.svg|thumb|255px|Magnetic field lines in a plane

bisecting the current loops. Note the field is approximately uniform in between the coil pair. (In this picture the coils are placed one beside the other: the axis is horizontal.)]]

thumb|255px|Magnetic field induction along the axis crossing the center of coils; z&nbsp;=&nbsp;0 is the point in the middle of the distance between coils

thumb|255px|Contours showing the magnitude of the magnetic field near a coil pair, with one coil at top and the other at bottom. Inside the central "octopus", the field is within 1% of its central value B<sub>0</sub>. The eight contours are for field magnitudes of 0.5&nbsp;B<sub>0</sub>, 0.8&nbsp;B<sub>0</sub>, 0.9&nbsp;B<sub>0</sub>, 0.95&nbsp;B<sub>0</sub>, 0.99&nbsp;B<sub>0</sub>, 1.01&nbsp;B<sub>0</sub>, 1.05&nbsp;B<sub>0</sub>, and 1.1&nbsp;B<sub>0</sub>.

The calculation of the exact magnetic field at any point in space is mathematically complex and involves the study of Bessel functions. Things are simpler along the axis of the coil-pair, and it is convenient to think about the Taylor series expansion of the field strength as a function of <math>x</math>, the distance from the central point of the coil-pair along the axis.

By symmetry, the odd-order terms in the expansion are zero. By arranging the coils so that the origin <math>x=0</math> is an inflection point for the field strength due to each coil separately, one can guarantee that the order <math>x^2</math> term is also zero, and hence the leading non-constant term is of order <math>x^4</math>. The inflection point for a simple coil is located along the coil axis at a distance <math>R/2</math> from its centre. Thus the locations for the two coils are <math>x=\pm R/2</math>.

The calculation detailed below gives the exact value of the magnetic field at the center point. If the radius is R, the number of turns in each coil is n and the current through the coils is I, then the magnetic field B at the midpoint between the coils will be given by

:<math> B = {\left ( \frac{4}{5} \right )}^{3/2} \frac{\mu_0 n I}{R},</math>

where <math>\mu_0</math> is the permeability of free space .

Derivation

Start with the formula for the on-axis field due to a single wire loop which is itself derived from the Biot–Savart law:

:<math> B_1(x) = \frac{\mu_0 I R^2}{2(R^2+x^2)^{3/2=\xi(x) \frac{\mu_0 I}{2R}.</math>

Here

:<math>\mu_0\;</math> = the permeability constant = <math> 4\pi \times 10^{-7} \text{ T}\cdot\text{m/A} = 1.257 \times 10^{-6} \text{ T}\cdot\text{m/A},</math>

:<math>I\;</math> = coil current, in amperes,

:<math>R\;</math> = coil radius, in meters,

:<math>x\;</math> = coil distance, on axis, to point, in meters,

:<math>\xi(x)=[1+(x/R)^2]^{-3/2}\;</math>is the distance dependent, dimensionless coefficient.

The Helmholtz coils consists of n turns of wire, so the equivalent current in a one-turn coil is n times the current I in the n-turn coil. Substituting nI for I in the above formula gives the field for an n-turn coil:

:<math> B_n(x) = \xi(x)\frac{\mu_0 n I}{2R}.</math>

Now provided that the coils are separated a distance <math>R</math> apart, we may already express the magnetic field right in-between the coils as <math>B_n(R/2)+B_n(-R/2)</math>, which produces the desired result. However, to justify why the coils ought to be separated by <math>R</math>, we need to do more work. We may, in general, set the coils a distance <math>a</math> apart, producing a field of magnitude <math>B_n(a/2)+B_n(-a/2)</math> at the center. However, this way of representing the solution carries no information about whether the field genuinely is nearly homogeneous between the coils. To understand the behavior of the field around the center point, we may define the function <math>B(x)=B_n(a/2+x)+B_n(-a/2+x)</math> to shift around the point of interest by <math>x</math>. If the magnetic field were truly homogeneous, all the derivatives of <math>B</math> with respect to <math>x</math> would vanish for all inputs <math>x</math>. There is no <math>a</math> for which this is precisely true, however, for the center point corresponding to <math>x=0</math> we can show that <math>a=R/2</math> is optimal in the sense of annihilating the first, second and third order derivatives.

The zeros of the derivatives of <math>B</math> are not affected by constant factors so it suffices to consider the derivatives of <math>\xi(a/2+x)+\xi(-a/2+x)</math>. To do this, we write the Taylor series, giving us

<blockquote><math>

\frac{2}{\left( 1+\frac{a^2}{R^2} \right)^{3/2+\frac{1}{2}\left( \frac{30a^2}{\left(1+\frac{a^2}{R^2}\right)^{7/2} R^4} -\frac{6}{\left(1+\frac{a^2}{R^2}\right)^{5/2} R^2}\right) x^2+\mathcal{O}(x^4).

</math></blockquote>

Since the coefficients of the Taylor series correspond to all the derivatives of the function, we notice that the first and the third derivatives of <math>B</math> are always zero at the center where <math>x=0</math> (due to the symmetry of the setup, true independently of the specific value of <math>a</math>) and for some special values of <math>a</math> the second order term may also vanish, corresponding to the derivative of the magnetic field <math>B</math> staying approximately zero around <math>x=0</math>, thus giving an approximately homogeneous field. The second order coefficient simplifies to

<blockquote><math>\frac{24a^2-6R^2}{(1+\frac{a^2}{R^2})^{7/2}R^4},</math></blockquote>

thus giving <math>a=R/2</math> as the desired parameter <math>a</math>. Therefore the coils ought to be a distance <math>R</math> apart, leading to an approximately homogeneous magnetic field inside the Helmholtz coil.

Time-varying magnetic field

Most Helmholtz coils use DC (direct) current to produce a static magnetic field. Many applications and experiments require a time-varying magnetic field. These applications include magnetic field susceptibility tests, scientific experiments, and biomedical studies (the interaction between magnetic field and living tissue). The required magnetic fields are usually either pulse or continuous sinewave. The magnetic field frequency range can be anywhere from near DC (0&nbsp;Hz) to many kilohertz or even megahertz (MHz). An AC Helmholtz coil driver is needed to generate the required time-varying magnetic field. The waveform amplifier driver must be able to output high AC current to produce the magnetic field.

Driver voltage and current

<math>I=\left ( \frac{5}{4} \right )^{3/2}\left ( \frac{BR}{\mu_0n} \right )</math>

Use the above equation in the mathematics section to calculate the coil current for a desired magnetic field, .

where <math>\mu_0</math> is the permeability of free space or <math> 4\pi \times 10^{-7} \text{ T}\cdot\text{m/A} = 1.257 \times 10^{-6} \text{ T}\cdot\text{m/A},</math>

<math>I\;</math> = coil current, in amperes,

<math>R\;</math> = coil radius, in meters,

n = number of turns in each coil.

thumb|304x304px|Using a function generator and a high-current waveform amplifier driver to generate high-frequency Helmholtz magnetic field

Then calculate the required Helmholtz coil driver amplifier voltage:

:<math>V=I\sqrt{\bigl[\omega\bigl(L_1+L_2\bigr)\bigr]^2+\bigl(R_1+R_2\bigr)^2}</math>

where

  • is the peak current,
  • is the angular frequency or ,
  • and are the inductances of the two Helmholtz coils, and
  • and are the resistances of the two coils.

High-frequency series resonant

Generating a static magnetic field is relatively easy; the strength of the field is proportional to the current. Generating a high-frequency magnetic field is more challenging. The coils are inductors, and their impedance increases proportionally with frequency. To provide the same field intensity at twice the frequency requires twice the voltage across the coil. Instead of directly driving the coil with a high voltage, a series resonant circuit may be used to provide the high voltage. A series capacitor is added in series with the coils. The capacitance is chosen to resonate the coil at the desired frequency. Only the coils parasitic resistance remains. This method only works at frequencies close to the resonant frequency; to generate the field at other frequencies requires different capacitors. The Helmholtz coil resonant frequency, <math>f_0</math>, and capacitor value, C, are given below.

  • Helmholtz coils were designed and built for the Army Research Laboratory's electromagnetic composite testing laboratory in 1993, for testing of composite materials to low-frequency magnetic fields.