A Geiger counter (, ; also known as a Geiger–Müller counter or G-M counter) is an electronic instrument for detecting and measuring ionizing radiation with the use of a Geiger–Müller tube. It is widely used in applications such as radiation dosimetry, radiological protection, experimental physics and the nuclear industry.
It detects ionizing radiation such as alpha particles, beta particles, and gamma rays using the ionization effect produced in a Geiger–Müller tube, which gives its name to the instrument. but it was not until the development of the Geiger–Müller tube in 1928 that the Geiger counter could be produced as a practical instrument. Since then, it has been very popular due to its robust sensing element and relatively low cost. However, there are limitations in measuring high radiation rates and the energy of incident radiation.
Principle of operation
thumb|upright=1.2|Diagram of a Geiger counter using an "end window" tube for low-penetration radiation. A loudspeaker is also used for indication.
A Geiger counter consists of a Geiger–Müller tube (the sensing element which detects the radiation) and the processing electronics, which display the result.
The Geiger–Müller tube is filled with an inert gas such as helium, neon, or argon at low pressure, to which a high voltage is applied. The tube briefly conducts electrical charge when high energy particles or gamma radiation make the gas conductive by ionization. The ionization is considerably amplified within the tube by the Townsend discharge effect to produce an easily measured detection pulse, which is fed to the processing and display electronics. This large pulse from the tube makes the Geiger counter relatively cheap to manufacture, as the subsequent electronics are greatly simplified. The correct voltage is usually specified by the manufacturer. To help quickly terminate each discharge in the tube a small amount of halogen gas or organic material known as a quenching mixture is added to the fill gas.
Readout
There are two types of detected radiation readout: counts and radiation dose.
- The counts display is the simplest, and shows the number of ionizing events detected, displayed either as a count rate, such as "counts per minute" or "counts per second", or as a total number of counts over a set time period (an integrated total). The counts readout is normally used when alpha or beta particles are being detected.
- More complex to achieve is a display of radiation dose rate, displayed in units such as the sievert, which is normally used for measuring gamma or X-ray dose rates. A Geiger–Müller tube can detect the presence of radiation, but not its energy, which influences the radiation's ionizing effect. Consequently, instruments measuring dose rate require the use of an energy compensated Geiger–Müller tube, so that the dose displayed relates to the counts detected. The electronics will apply known factors to make this conversion, which is specific to each instrument and is determined by design and calibration.
The readout can be analog or digital, and modern instruments offer serial communications with a host computer or network.
There is usually an option to produce audible clicks representing the number of ionization events detected. This is the distinctive sound associated with handheld or portable Geiger counters. The purpose of this is to allow the user to concentrate on manipulation of the instrument while retaining auditory feedback on the radiation rate.
Limitations
There are two main limitations of the Geiger counter:
- Because the output pulse from a Geiger–Müller tube is always of the same magnitude (regardless of the energy of the incident radiation), the tube cannot differentiate between radiation types or measure radiation energy, which prevents it from correctly measuring dose rate. However, for discrimination between α- and β-particles or provision of particle energy information, scintillation counters or proportional counters should be used.
Neutron detection
lang=en|thumb|Geiger tube filled with BF<sub>3</sub> for detection of thermal neutrons
A variation of the Geiger tube can be used to exclusively measure radiation dosage from neutrons rather than from gammas by the process of neutron capture. The tube, which can contain the fill gas helium-3, is surrounded by a plastic moderator that reduces neutron energies prior to capture. When a capture occurs in the fill gas, the energy released is registered in the detector.
thumb|A modern one-piece Geiger–Müller counter, including Geiger–Müller tube type 70 019 (at the top)
Gamma measurement—personnel protection and process control
While "Geiger counter" is practically synonymous with the hand-held variety, the Geiger principle is in wide use in installed "area gamma" alarms for personnel protection, as well as in process measurement and interlock applications. The processing electronics of such installations have a higher degree of sophistication and reliability than those of hand-held meters.
Physical design
thumb|right|Pancake G-M tube used for alpha and beta detection; the delicate mica window is usually protected by a mesh when fitted in an instrument.
For hand-held units there are two fundamental physical configurations: the "integral" unit with both detector and electronics in the same unit, and the "two-piece" design which has a separate detector probe and an electronics module connected by a short cable.
In the 1930s a mica window was added to the cylindrical design allowing low-penetration radiation to pass through with ease.
The integral unit allows single-handed operation, so the operator can use the other hand for personal security in challenging monitoring positions, but the two piece design allows easier manipulation of the detector, and is commonly used for alpha and beta surface contamination monitoring where careful manipulation of the probe is required or the weight of the electronics module would make operation unwieldy. A number of different sized detectors are available to suit particular situations, such as placing the probe in small apertures or confined spaces.
Gamma and X-Ray detectors generally use an "integral" design so the Geiger–Müller tube is conveniently within the electronics enclosure. This can easily be achieved because the casing usually has little attenuation, and is employed in ambient gamma measurements where distance from the source of radiation is not a significant factor. However, to facilitate more localised measurements such as "surface dose", the position of the tube in the enclosure is sometimes indicated by targets on the enclosure so an accurate measurement can be made with the tube at the correct orientation and a known distance from the surface.
There is a particular type of gamma instrument known as a "hot spot" detector which has the detector tube on the end of a long pole or flexible conduit. These are used to measure high radiation gamma locations whilst protecting the operator by means of distance shielding.
Particle detection of alpha and beta can be used in both integral and two-piece designs. A pancake probe (for alpha/beta) is generally used to increase the area of detection in two-piece instruments whilst being relatively light weight. In integral instruments using an end window tube there is a window in the body of the casing to prevent shielding of particles. There are also hybrid instruments which have a separate probe for particle detection and a gamma detection tube within the electronics module. The detectors are switchable by the operator, depending the radiation type that is being measured.
Guidance on application use
In the United Kingdom the National Radiological Protection Board issued a user guidance note on selecting the best portable instrument type for the radiation measurement application concerned. This covers all radiation protection instrument technologies and includes a guide to the use of G-M detectors.
History
thumb|right|Rutherford & Geiger particle counter (1908). Alpha particles from a source in the firing tube were admitted through aperture "D" to the detecting vessel, which was a brass tube with a central co-axial wire "B" at a relative potential of 1320 volts DC. The aperture had a mica window so the detecting vessel could run at an [[absolute pressure of 2-5 mm of mercury. These conditions enabled creation of a Townsend discharge for every alpha particle entering the vessel. At very low count rates these registered as "kicks" on an electrometer needle.]]
thumb|Early Geiger–Müller tube made in 1932 by Hans Geiger for laboratory use
In 1908 Hans Geiger, under the supervision of Ernest Rutherford at the Victoria University of Manchester (now the University of Manchester), developed an experimental technique for detecting alpha particles that would later be used to develop the Geiger–Müller tube in 1928. This early counter was only capable of detecting alpha particles and was part of a larger experimental apparatus. The fundamental ionization mechanism used was discovered by John Sealy Townsend between 1897 and 1901, and is known as the Townsend discharge, which is the ionization of molecules by ion impact.
It was not until 1928 that Geiger and Walther Müller (a PhD student of Geiger) developed the sealed Geiger–Müller tube which used basic ionization principles previously used experimentally. Small and rugged, not only could it detect alpha and beta radiation as prior models had done, but also gamma radiation. Now a practical radiation instrument could be produced relatively cheaply, and so the Geiger counter was born. As the tube output required little electronic processing, a distinct advantage in the thermionic valve era due to minimal valve count and low power consumption, the instrument achieved great popularity as a portable radiation detector.
Modern versions of the Geiger counter use halogen quench gases, a technique invented in 1947 by Sidney H. Liebson. Halogen compounds have superseded the organic quench gases because of their much longer life and lower operating voltages; typically 400-900 volts.
Gallery
<gallery>
File:Transuranic waste casks.jpg|Use of a "hot spot" detector on a long pole to survey waste casks.
File:RM-80 GM with LCD-90 Micro Controller and Wireless Bluetooth.jpg|G-M pancake detector (right) feeding a microcontroller data-logger (left) sending data to a PC via bluetooth. A radioactive rock was placed on the detector causing the graph (in background) to rise.
File:Cosmos 954 - Recovery 001.jpg|G-M counters being used in the search for debris of satellite Kosmos 954
</gallery>
See also
- Becquerel, the SI unit of the radioactive decay rate of a quantity of radioactive material
- Civil defense Geiger counters, handheld radiation monitors, both G-M and ion chambers
- Counting efficiency the ratio of radiation events reaching a detector and the number it counts
- Data sonification, the interpretation or processing of data by sound
- Dosimeter, a device used by personnel to measure what radiation dose they have received
- Ionization chamber, the simplest ionising radiation detector
- Gaseous ionization detector, an overview of the main gaseous detector types
- Geiger–Müller tube, provides a more detailed description of Geiger–Müller tube operation and types
- Geiger plateau, the correct operating voltage range for a Geiger–Müller tube
- Photon counting
- Radioactive decay, the process by which unstable atoms emit radiation
- Safecast (organization), use of Geiger–Müller counter technology in citizen science
- Scintillation counter, a gasless radiation detector
- Sievert, the SI unit of stochastic effects of radiation on the human body